Measurement of the 226Th and 222Ra half-lives

Appl Radiat Isot. 2012 Sep;70(9):1913-8. doi: 10.1016/j.apradiso.2012.02.117. Epub 2012 Mar 7.

Abstract

The half-lives of (226)Th and (222)Ra were measured by counting alpha-particle emissions from sources as a function of time. The (226)Th sources were prepared from an open (230)U source, capturing recoil atoms after alpha-particle decay on glass disks or even directly onto a detector. Similarly, the (222)Ra sources were obtained by self-transfer of recoil atoms from (226)Th sources. The activity measurements were performed in (nearly-)2π geometry with an ion-implanted silicon detector. The decay curves were analysed in different ways, incl. moment analysis, trying to avoid the pitfalls of bias of least-squares fits to Poisson distributed data. The observed half-life values are T(1/2)((226)Th)=30.70 (3) min and T(1/2)((222)Ra)=33.6 (4) s. Literature values show some inconsistency.

Publication types

  • Comparative Study

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis
  • Half-Life
  • Radiation Dosage
  • Radiometry / instrumentation*
  • Radium / analysis
  • Radium / chemistry*
  • Thorium / analysis
  • Thorium / chemistry*

Substances

  • Thorium
  • Radium