[Degenerate oligonucleotide primed-PCR technology establishment and its sensitivity test analysis]

Fa Yi Xue Za Zhi. 2012 Feb;28(1):41-3.
[Article in Chinese]

Abstract

Objective: To establish a whole genome amplification testing system based on degenerate oligonucleotide primed-PCR (DOP-PCR) and to explore its reliability and sensitivity.

Methods: DOP-PCR amplified production was detected by fluorescent labeled multiplex STR amplification and capillary electrophoresis detection system to determine reliability and sensitivity of DOP-PCR system.

Results: DOP-PCR system was successfully established and the detection sensitivity reached 5 cells (30 pg) by pretreatment of DOP-PCR and then detection of STR genotyping.

Conclusion: The system established in this study is reliable and more testing sensitive for forensic trace evidence.

Publication types

  • English Abstract
  • Research Support, Non-U.S. Gov't

MeSH terms

  • DNA / analysis*
  • DNA / genetics*
  • DNA Fingerprinting / methods
  • DNA Primers*
  • Electrophoresis, Capillary
  • Gene Amplification
  • Genotype
  • Humans
  • Nucleic Acid Amplification Techniques
  • Oligonucleotides / genetics
  • Polymerase Chain Reaction / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tandem Repeat Sequences*

Substances

  • DNA Primers
  • Oligonucleotides
  • DNA