Standardization of 64Cu and 68Ga by the 4π(PC)β-γ coincidence method and calibration of the ionization chamber

Appl Radiat Isot. 2012 Sep;70(9):2025-30. doi: 10.1016/j.apradiso.2012.02.043. Epub 2012 Mar 3.

Abstract

The paper treats the application of the 4π(PC)β-γ coincidence method for the standardization of the radionuclides (64)Cu and (68)Ga. The general coincidence equations are written. Two types of extrapolation were described and used in measurement: the positron-annihilation coincidence, and the counting of all emitted radiations; both methods are compared with respect to results, advantages and drawbacks. The impurities' content correction was applied. The standardized solutions were used to calibrate the ionization chamber CENTRONIC IG12/20A and to determine the gamma-rays emission intensities.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Calibration
  • Copper Radioisotopes / analysis
  • Copper Radioisotopes / chemistry*
  • Copper Radioisotopes / standards*
  • Equipment Design
  • Equipment Failure Analysis
  • Gallium Radioisotopes / analysis
  • Gallium Radioisotopes / chemistry*
  • Gallium Radioisotopes / standards*
  • Half-Life
  • Internationality
  • Radiation Dosage
  • Radiometry / instrumentation*
  • Radiometry / standards*
  • Reference Standards
  • Reference Values