Reduced-distortion diffusion MRI of the craniovertebral junction

AJNR Am J Neuroradiol. 2012 Aug;33(7):1321-5. doi: 10.3174/ajnr.A2969. Epub 2012 Mar 1.

Abstract

Background and purpose: CVJ lesion suffers from a high sensitivity to susceptibility and distortion artifacts, which sometimes makes diffusion image difficult to interpret. Our purpose was to evaluate the potential for diffusion MR imaging using RS-EPI compared with SS-EPI in the assessment of the CVJ.

Materials and methods: RS-EPI and SS-EPI DTI images were acquired from 10 healthy volunteers using 3T MRI with a 32-channel head coil. For both sequences, the following parameters were used: 1-mm(2) in-plane resolution; 3-mm section thickness; TR = 5200 ms; 1 acquisition at b = 0 and 12 different encoding directions at b = 1000 seconds/mm(2). The RS-EPI sequence scan time was 9.44 minutes (1 average). The SS-EPI sequence was 9.37 minutes (8 averages). Diffusion tensor calculation and image analysis were performed using DTIStudio software. Diffusion trace images and color-coded fiber orientation maps were evaluated by 2 independent readers for distortion and delineation of fine structure using a semiquantitative scale in selected landmark locations. The absolute distances between the temporal base and the cerebellar contour between the T2-weighted images and the diffusion trace images obtained with RS-EPI and SS-EPI were also compared.

Results: The contours of the temporal lobe and cerebellum were better delineated and distortion artifacts were clearly reduced with the RS-EPI sequence. More fine structures were also visible in the brain stem and cerebellum with the RS-EPI sequence. The amount of distortion was significantly reduced with RS-EPI compared with SS-EPI (P < .01).

Conclusions: The RS-EPI DTI sequence was less prone to geometric distortion than the SS-EPI sequence and allowed a better delineation of CVJ internal structure. Although the acquisition time is still relatively long, the RS-EPI appears as a promising approach to perform DTI studies in CVJ lesions, such as brain stem ischemia, neurodegenerative diseases, brain and skull base tumors, or inflammation.

MeSH terms

  • Adult
  • Algorithms
  • Artifacts*
  • Cervical Atlas / anatomy & histology*
  • Diffusion Magnetic Resonance Imaging / methods*
  • Female
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Ligaments / anatomy & histology*
  • Male
  • Models, Biological
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Skull Base / anatomy & histology*