A disposable chitosan-modified carbon fiber electrode for dengue virus envelope protein detection

Talanta. 2012 Mar 15:91:41-6. doi: 10.1016/j.talanta.2012.01.002. Epub 2012 Jan 5.

Abstract

A chitosan-modified carbon fiber electrode (CFE) for dengue virus envelope protein (DENV) was developed. Antibodies against DENV were covalently immobilized on the chitosan (CHIT) matrix after activation with sodium periodate. Cyclic voltammetries and scanning electron microscopies analysis were performed to monitor steps involved in the CFE surface modification. Amperometric response of the competitive immunoassays was generated by hydrogen peroxide reaction with the peroxidase conjugated to DENV and 2'-azino-bis-(3-ethylbenzthiazoline-6-sulfonic acid) (ABTS) as mediator. The immunosensor showed a lower limit of detection for DENV (0.94 ng mL(-1)) than previously described and a linear range from 1.0 to 175 ng mL(-1), in concentration levels clinically relevant for dengue virus diagnosis. The intra- and inter-assay were respectively 5.8% and 3.6%. The unique and simple design of this immunoassay format provides an economical alternative for the manufacture of other sensitive sensors.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Carbon
  • Carbon Fiber
  • Chitosan
  • Dengue / diagnosis
  • Dengue Virus / immunology
  • Dengue Virus / isolation & purification*
  • Electrodes / standards*
  • Immunoassay / methods
  • Immunoassay / standards
  • Viral Envelope Proteins / analysis*
  • Viral Envelope Proteins / immunology

Substances

  • Carbon Fiber
  • Viral Envelope Proteins
  • Carbon
  • Chitosan