Video-coaching as biofeedback tool to improve gated treatments: Possibilities and limitations

Z Med Phys. 2012 Sep;22(3):224-30. doi: 10.1016/j.zemedi.2012.01.001. Epub 2012 Feb 20.

Abstract

For respiratory gated radiotherapy the manufacturers of linear accelerators offer dedicated gating technologies. The video-based Varian RPM Gating system (Varian Medical Systems, Palo Alto/CA, USA) includes in a standard configuration a support tool for regular breathing called audiocoaching. As this approach has limitations regarding direct control of the patient's breathing due to a missing feedback, we designed an additional tool offering videocoaching. In order to evaluate the impact of this additional functionality, we measured parameters defining the image quality of 4D-CT data as well as the treatment duration which is mainly influenced by the patient's limited ability to achieve a stable breathing pattern.

MeSH terms

  • Biofeedback, Psychology / methods*
  • Breath Holding*
  • Humans
  • Radiographic Image Enhancement / methods*
  • Radiotherapy, Conformal / methods
  • Radiotherapy, Image-Guided / methods*
  • Reproducibility of Results
  • Respiratory-Gated Imaging Techniques / methods*
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / methods*
  • Video Recording / methods*