Flip angle profile correction for T₁ and T₂ quantification with look-locker inversion recovery 2D steady-state free precession imaging

Magn Reson Med. 2012 Nov;68(5):1579-85. doi: 10.1002/mrm.24151. Epub 2012 Jan 31.

Abstract

Fast methods using balanced steady-state free precession have been developed to reduce the scan time of T₁ and T₂ mapping. However, flip angle (FA) profiles created by the short radiofrequency pulses used in steady-state free precession deviate substantially from the ideal rectangular profile, causing T₁ and T₂ mapping errors. The purpose of this study was to develop a FA profile correction for T₁ and T₂ mapping with Look-Locker 2D inversion recovery steady-state free precession and to validate this method using 2D spin echo as a reference standard. Phantom studies showed consistent improvement in T₁ and T₂ accuracy using profile correction at multiple FAs. Over six human calves, profile correction provided muscle T₁ estimates with mean error ranging from excellent (-0.6%) at repetition time/FA = 18 ms/60° to acceptable (6.8%) at repetition time/FA = 4.9 ms/30°, while muscle T₂ estimates were less accurate with mean errors of 31.2% and 47.9%, respectively.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Algorithms*
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Muscle, Skeletal / anatomy & histology*
  • Reproducibility of Results
  • Sensitivity and Specificity