Stacking open-capillary electroosmotic pumps in series to boost the pumping pressure to drive high-performance liquid chromatographic separations

J Chromatogr A. 2012 Mar 2:1227:253-8. doi: 10.1016/j.chroma.2011.12.105. Epub 2012 Jan 9.

Abstract

Numerous micropumps have been developed, but few of them can produce adequate flow rate and pressure for high-performance liquid chromatography (HPLC) applications. We have recently developed an innovative hybrid electroosmotic pump (EOP) to solve this problem. The basic unit of a hybrid pump consists of a +EOP (the pumping element is positively charged) and a -EOP (the pumping element is negatively charged). The outlet of the +EOP is then joined with the inlet of the -EOP, forming a basic pump unit, while the anode of a positive high voltage (HV) power supply is placed at the joint. The inlet and outlet of this pump unit are electrically grounded. With this configuration, we can stack many of such pump units in series to boost the pumping power. In this work, we describe in details how an open-capillary hybrid EOP is constructed and characterize this pump systematically. We also show that a hybrid EOP with ten serially stacked pump units can deliver a maximum pressure of 21.5 MPa (∼3100 psi). We further demonstrate the feasibility of using this hybrid EOP to drive eluents for HPLC separations of proteins and peptides.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Acetonitriles / chemistry
  • Chromatography, High Pressure Liquid / instrumentation*
  • Chromatography, High Pressure Liquid / methods*
  • Electroosmosis / instrumentation*
  • Peptides / isolation & purification
  • Pressure
  • Proteins / isolation & purification
  • Trifluoroacetic Acid / chemistry

Substances

  • Acetonitriles
  • Peptides
  • Proteins
  • Trifluoroacetic Acid
  • acetonitrile