Transport of redox probes through single pores measured by scanning electrochemical-scanning ion conductance microscopy (SECM-SICM)

Analyst. 2012 Jul 7;137(13):2933-8. doi: 10.1039/c2an16178h. Epub 2012 Jan 26.

Abstract

We report scanning electrochemical microscopy-scanning ion conductance microscopy (SECM-SICM) experiments that describe transport of redox active molecules which emanate from single pores of a track-etch membrane. Experiments are performed with electrodes which consist of a thin gold layer deposited on one side of a nanopipet. Subsequent insulation of the electrode with parylene results in a hybrid electrode for SECM-SICM measurements. Electrode fabrication is straightforward and highly parallel. For image collection, ionic current measured at the nanopipet both controls the position of the electrode with respect to the membrane surface and reports the local conductance in the vicinity of the nanopipet, while faradaic current measured at the Au electrode reports the presence of redox-active molecules. Application of a transmembrane potential difference affords additional control over migration of charged species across the membrane.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electrochemical Techniques / methods*
  • Electrodes
  • Microscopy / methods*
  • Oxidation-Reduction