Evaluation of the analytical performance of the coulometry-based Optium Omega blood glucose meter

J Diabetes Sci Technol. 2011 Nov 1;5(6):1612-7. doi: 10.1177/193229681100500640.

Abstract

Background: The goal of diabetes treatment is maintaining near normoglycemia based on self-monitoring of blood glucose (SMBG). In this study, an evaluation of the analytical performance of the coulometry-based Optium Omega™ glucose meter designed for SMBG has been carried out.

Methods: The assessment of precision and between-lot variability was based on glucose measurements in ethylene-diaminetetraacetic acid venous blood samples. Glucose concentrations measured in 289 fresh capillary blood samples using the Omega glucose meter and the Biosen C_line analyzer were compared.

Results: Within-run imprecision coefficient of variation for the lower and higher glucose concentrations amounted to 5.09 and 2.1%, respectively. The relative lot-dependent differences found for the lower and higher glucose concentrations were equal to 6.8 and 2.6%, respectively. The glucose meter error calculated for various concentration ranges amounted from 2.22 to 4.48%. The glucose meter error met the accuracy criteria recommended by the International Organization for Standardization and the American Diabetes Association. The Passing-Bablok agreement test and error grid analysis with 96% of results in zone A indicated good concordance of results, including glucose concentrations below 100 mg/dl.

Conclusions: The evaluated Optium Omega glucose meter fits the analytical requirements for its use in blood glucose monitoring in diabetes patients.

Publication types

  • Evaluation Study

MeSH terms

  • Blood Chemical Analysis / methods*
  • Blood Glucose / analysis*
  • Diabetes Mellitus / blood*
  • Electrolysis / methods*
  • Humans
  • Reproducibility of Results

Substances

  • Blood Glucose