Thermal diffusivity imaging with the thermal lens microscope

Appl Opt. 2011 Dec 1;50(34):6336-42. doi: 10.1364/AO.50.006336.

Abstract

A coaxial thermal lens microscope was used to generate images based on both the absorbance and thermal diffusivity of histological samples. A pump beam was modulated at frequencies ranging from 50 kHz to 5 MHz using an acousto-optic modulator. The pump and a CW probe beam were combined with a dichroic mirror, directed into an inverted microscope, and focused onto the specimen. The change in the transmitted probe beam's center intensity was detected with a photodiode. The photodiode's signal and a reference signal from the modulator were sent to a high-speed lock-in amplifier. The in-phase and quadrature signals were recorded as a sample was translated through the focused beams and used to generate images based on the amplitude and phase of the lock-in amplifier's signal. The amplitude is related to the absorbance and the phase is related to the thermal diffusivity of the sample. Thin sections of stained liver and bone tissues were imaged; the contrast and signal-to-noise ratio of the phase image was highest at frequencies from 0.1-1 MHz and dropped at higher frequencies. The spatial resolution was 2.5 μm for both amplitude and phase images, limited by the pump beam spot size.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Bone and Bones / ultrastructure
  • Equipment Design
  • Lasers
  • Lenses*
  • Liver / ultrastructure
  • Microscopy / instrumentation*
  • Microscopy / methods*
  • Signal-To-Noise Ratio
  • Temperature