Assessment of radiant temperature in a closed incubator

Eur J Appl Physiol. 2012 Aug;112(8):2957-68. doi: 10.1007/s00421-011-2265-9. Epub 2011 Dec 10.

Abstract

In closed incubators, radiative heat loss (R) which is assessed from the mean radiant temperature (Tr) accounts for 40-60% of the neonate's total heat loss. In the absence of a benchmark method to calculate Tr--often considered to be the same as the air incubator temperature-errors could have a considerable impact on the thermal management of neonates. We compared Tr using two conventional methods (measurement with a black-globe thermometer and a radiative "view factor" approach) and two methods based on nude thermal manikins (a simple, schematic design from Wheldon and a multisegment, anthropometric device developed in our laboratory). By taking the Tr estimations for each method, we calculated metabolic heat production values by partitional calorimetry and then compared them with the values calculated from V(O2) and V(CO2) measured in 13 preterm neonates. Comparisons between the calculated and measured metabolic heat production values showed that the two conventional methods and Wheldon's manikin underestimated R, whereas when using the anthropomorphic thermal manikin, the simulated versus clinical difference was not statistically significant. In conclusion, there is a need for a safety standard for measuring TR in a closed incubator. This standard should also make available estimating equations for all avenues of the neonate's heat exchange considering the metabolic heat production and the modifying influence of the thermal insulation provided by the diaper and by the mattress. Although thermal manikins appear to be particularly appropriate for measuring Tr, the current lack of standardized procedures limits their widespread use.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Body Temperature Regulation*
  • Energy Metabolism
  • Environment, Controlled
  • Equipment Design
  • France
  • Gestational Age
  • Humans
  • Incubators, Infant* / standards
  • Infant
  • Infant, Low Birth Weight
  • Infant, Newborn
  • Infant, Premature*
  • Infant, Very Low Birth Weight
  • Manikins
  • Materials Testing
  • Models, Biological
  • Oxygen Consumption
  • Skin Temperature
  • Supine Position
  • Temperature*
  • Thermogenesis
  • Thermography* / standards
  • Thermometers