Time-optimized X-ray micro-CT imaging of polymer based scaffolds

J Biomed Mater Res B Appl Biomater. 2012 Feb;100(2):360-7. doi: 10.1002/jbm.b.31957. Epub 2011 Nov 24.

Abstract

Although X-ray microscopic computed tomography is widely used to assess the structural properties of polymeric tissue scaffolds its validity is dependent on the quality of the images obtained. Here, the role of resolution, integration time, image averaging, and X-ray power on the accurate determination of scaffold porosity, while aiming to minimize imaging time, was investigated. This work identified key parameters for optimization and a methodology to vary them to improve results. Based on this, guidelines were developed to assist in the selection of image acquisition parameters to allow rapid and accurate scaffold imaging as required for mass manufacture.

Keywords: image analysis; micro-CT; porosity; regenerative medicine; scaffold characterization.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Polymers / chemistry
  • Tissue Scaffolds*
  • X-Ray Microtomography / methods*

Substances

  • Polymers