Noise immune cavity enhanced optical heterodyne velocity modulation spectroscopy

Opt Express. 2011 Nov 21;19(24):24822-7. doi: 10.1364/OE.19.024822.

Abstract

The novel technique of cavity enhanced velocity modulation spectroscopy has recently been demonstrated as the first general absorption technique that allows for sub-Doppler spectroscopy of molecular ions while retaining ion-neutral discrimination. The previous experimental setup has been further improved with the addition of heterodyne detection in a NICE-OHMS setup. This improves the sensitivity by a factor of 50 while retaining sub-Doppler resolution and ion-neutral discrimination. Calibration was done with an optical frequency comb, and line centers for several N(2)(+) lines have been determined to within an accuracy of 300 kHz.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis
  • Optical Devices*
  • Refractometry / instrumentation*
  • Spectrum Analysis / instrumentation*
  • Surface Plasmon Resonance / instrumentation*