Sensitive biosensors using Fano resonance in single gold nanoslit with periodic grooves

Opt Express. 2011 Nov 21;19(24):24530-9. doi: 10.1364/OE.19.024530.

Abstract

Chip-based biosensors for sensitive label-free detection were fabricated and tested by using Fano-type resonant nanostructures. The sensor was composed of a 190 nm-thick gold nanoslit surrounded by 600-nm-period grooves. Transverse-magnetic polarized wave in these gold nanostructures generated asymmetrical resonant spectra due to the interference of broad-band cavity resonance in the single slit and narrow-band surface plasmon resonance on the periodic grooves. Compared to nanoslit arrays, such Fano-type sensor has a sharper resonance which yields a figure of merit up to 48. In addition, the crossed talk between sensing elements is reduced due to the Bragg reflection of the periodic grooves. A smaller detection separation down to 10 μm width was achieved. An antigen-antibody interaction experiment in aqueous environment verified the detection sensitivity in surface binding event.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biosensing Techniques / instrumentation*
  • Equipment Design
  • Equipment Failure Analysis
  • Gold / chemistry*
  • Immunoassay / instrumentation*
  • Nanostructures / chemistry*
  • Nanostructures / ultrastructure
  • Nanotechnology / instrumentation*
  • Surface Plasmon Resonance / instrumentation*

Substances

  • Gold