Comparison of real-time three-dimensional echocardiography with cardiovascular magnetic resonance for left ventricular volumetric assessment in unselected patients

Eur Heart J Cardiovasc Imaging. 2012 Feb;13(2):187-95. doi: 10.1093/ejechocard/jer248. Epub 2011 Nov 22.

Abstract

Aims: To compare left ventricular (LV) volume indices and the ejection fraction (EF) obtained using real-time three-dimensional echocardiography (RT3DE) and cardiovascular magnetic resonance (CMR) in unselected patients representative of 'real-world' clinical practice, and to determine the effect of RT3DE image quality on these parameters.

Methods and results: Sixty consecutive patients undergoing CMR underwent same day RT3DE. LV volume and EF measurements were made using both modalities and compared. All scans were independently analysed by a second observer to assess inter-observer variability, and 40% were re-analysed to assess intra-observer variability. RT3DE image quality was graded as good, adequate, and non-analysable. Thirteen (22%) patients had good RT3DE image quality, 29 (48%) had adequate image quality, and 18 (30%) had image quality precluding analysis. Body mass index and arrhythmia frequency were higher in patients with suboptimal image quality. RT3DE significantly underestimated end-diastolic volume (EDV) (-45 ± 35 mL, P < 0.001), end-systolic volume (ESV) (-11 ± 24 mL, P = 0.004), and EF (-7 ± 9%, P < 0.001) compared with CMR although the degree of underestimation was substantially less when image quality was good. Eleven patients (18%) classified as having a normal EF by CMR had a reduced EF according to RT3DE, all but one of which had suboptimal image quality. Observer variability for RT3DE was higher than for CMR for all parameters, however, the difference was not significant when RT3DE image quality was good.

Conclusions: In contrast to previously published data from highly selected patient groups, 'real-world' RT3DE substantially underestimates LV volumes and EF. The degree of underestimation is related to image quality.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Echocardiography, Three-Dimensional* / methods
  • Female
  • Heart Ventricles / diagnostic imaging*
  • Heart Ventricles / pathology*
  • Humans
  • Magnetic Resonance Imaging*
  • Male
  • Middle Aged
  • Observer Variation
  • Predictive Value of Tests
  • Prospective Studies
  • Sensitivity and Specificity
  • Stroke Volume*
  • Time Factors