[Determination of diborane in the air of workplace by ICP-AES]

Zhonghua Lao Dong Wei Sheng Zhi Ye Bing Za Zhi. 2011 Jun;29(6):452-4.
[Article in Chinese]

Abstract

Objective: A sampling method was established to collect diborane in the air of workplace and an ICP-AES method was developed to determine the Boron in desorbed solution.

Method: Diborane in the air of workplace was collected by solid sorbent tube filled with oxidant impregnated activated carbon. The adsorbed diborane was desorbed into 3% H2O2 aqueous, and then the desorbed Boron was determined by ICP-AES.

Results: The sampling efficiency of this method was 99.6% with the desorption efficiency of diborane with 5.660 microg and 56.6 microg spiked were 90.9% and 99.5%, respectively. Both the intra-and inter-precision RSD were less than 8%. The standard curve of this method ranged from 0.1 to 10.0 microg/ml (Boron), and the LOD and LOQ were 0.011 mg/m3 and 0.035 mg/m3 (15L samples) respectively.

Conclusion: The method established was suitable for diborane sampling and determination in the air of workplace.

Publication types

  • English Abstract
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Air Pollutants, Occupational / analysis*
  • Borohydrides / analysis*
  • Environmental Monitoring / methods
  • Spectrophotometry, Atomic / methods*
  • Workplace

Substances

  • Air Pollutants, Occupational
  • Borohydrides
  • diborane