Surface acoustic wave induced particle manipulation in a PDMS channel--principle concepts for continuous flow applications

Biomed Microdevices. 2012 Apr;14(2):279-89. doi: 10.1007/s10544-011-9606-7.

Abstract

A device for acoustic particle manipulation in the 40 MHz range for continuous-flow operation in a 50 μm wide PDMS channel has been evaluated. Unidirectional interdigital transducers on a Y-cut Z-propagation lithium nixobate wafer were used to excite a surface acoustic wave that generated an acoustic standing wave inside the microfluidic channel. It was shown that particle alignment nodes with different inter-node spacing could be obtained, depending on device design and driving frequency. The observed inter-node spacing differed from the standard half-wavelength inter-node spacing generally employed in bulk acoustic transducer excited resonant systems. This effect and the related issue of acoustic node positions relative the channel walls, which is fundamental for most continuous flow particle manipulation operations in channels, was evaluated in measurements and simulations. Specific applications of particle separation and alignment where these systems can offer benefits relative state-of the art designs were identified.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acoustics / instrumentation*
  • Cell Survival
  • Equipment Design
  • Microfluidic Analytical Techniques / instrumentation*
  • Microfluidic Analytical Techniques / methods*
  • Microfluidics
  • Sound*
  • Transducers
  • Ultrasonics