Low-cost adaptive astigmatism compensator for improvement of eye fundus camera

Opt Lett. 2011 Nov 1;36(21):4164-6. doi: 10.1364/OL.36.004164.

Abstract

In this work we show a conceptually simple and cheap means by which to improve retinal image quality in fundus cameras. We will explain how to build a low-cost variable astigmatism-correcting device with a pair of identical commercial ophthalmic lenses. We have developed and tested a low order aberration compensation device capable of correcting astigmatism prescriptions up to -8.00 D with no significant addition of higher order aberrations. A theoretical description of the device, calibration, and improvement in retinal images without employing any image restoration technique will be shown.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Astigmatism / pathology
  • Eye / anatomy & histology*
  • Humans
  • Lenses / economics
  • Optical Phenomena
  • Retina / anatomy & histology
  • Retinoscopes* / economics