Direct liquid chromatography method for the simultaneous quantification of hydroxytyrosol and tyrosol in red wines

J Agric Food Chem. 2011 Nov 9;59(21):11683-9. doi: 10.1021/jf202254t. Epub 2011 Oct 5.

Abstract

A validated HPLC method with fluorescence detection for the simultaneous quantification of hydroxytyrosol and tyrosol in red wines is described. Detection conditions for both compounds were optimized (excitation at 279 and 278 and emission at 631 and 598 nm for hydroxytyrosol and tyrosol, respectively). The validation of the analytical method was based on selectivity, linearity, robustness, detection and quantification limits, repeatability, and recovery. The detection and quantification limits in red wines were set at 0.023 and 0.076 mg L(-1) for hydroxytyrosol and at 0.007 and 0.024 mg L(-1) for tyrosol determination, respectively. Precision values, both within-day and between-day (n = 5), remained below 3% for both compounds. In addition, a fractional factorial experimental design was developed to analyze the influence of six different conditions on analysis. The final optimized HPLC-fluorescence method allowed the analysis of 30 nonpretreated Spanish red wines to evaluate their hydroxytyrosol and tyrosol contents.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Phenylethyl Alcohol / analogs & derivatives*
  • Phenylethyl Alcohol / analysis
  • Wine / analysis*

Substances

  • 3,4-dihydroxyphenylethanol
  • 4-hydroxyphenylethanol
  • Phenylethyl Alcohol