Automatic computed tomography patient dose calculation using DICOM header metadata

Radiat Prot Dosimetry. 2011 Sep;147(1-2):317-20. doi: 10.1093/rpd/ncr338. Epub 2011 Aug 9.

Abstract

The present work describes a method that calculates the patient dose values in computed tomography (CT) based on metadata contained in DICOM images in support of patient dose studies. The DICOM metadata is preprocessed to extract necessary calculation parameters. Vendor-specific DICOM header information is harmonized using vendor translation tables and unavailable DICOM tags can be completed with a graphical user interface. CT-Expo, an MS Excel application for calculating the radiation dose, is used to calculate the patient doses. All relevant data and calculation results are stored for further analysis in a relational database. Final results are compiled by utilizing data mining tools. This solution was successfully used for the 2009 CT dose study in Luxembourg. National diagnostic reference levels for standard examinations were calculated based on each of the countries' hospitals. The benefits using this new automatic system saved time as well as resources during the data acquisition and the evaluation when compared with earlier questionnaire-based surveys.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Automation
  • Body Burden
  • Humans
  • Image Interpretation, Computer-Assisted*
  • Radiation Dosage*
  • Radiology Information Systems*
  • Radiometry
  • Tomography, X-Ray Computed*