Effect of pulsatility on the mathematical modeling of rotary blood pumps

Artif Organs. 2011 Aug;35(8):825-32. doi: 10.1111/j.1525-1594.2011.01276.x. Epub 2011 Jul 27.

Abstract

In this study, the effect of time derivatives of flow rate and rotational speed was investigated on the mathematical modeling of a rotary blood pump (RBP). The basic model estimates the pressure head of the pump as a dependent variable using measured flow and speed as predictive variables. Performance of the model was evaluated by adding time derivative terms for flow and speed. First, to create a realistic working condition, the Levitronix CentriMag RBP was implanted in a sheep. All parameters from the model were physically measured and digitally acquired over a wide range of conditions, including pulsatile speed. Second, a statistical analysis of the different variables (flow, speed, and their time derivatives) based on multiple regression analysis was performed to determine the significant variables for pressure head estimation. Finally, different mathematical models were used to show the effect of time derivative terms on the performance of the models. In order to evaluate how well the estimated pressure head using different models fits the measured pressure head, root mean square error and correlation coefficient were used. The results indicate that inclusion of time derivatives of flow and speed can improve model accuracy, but only minimally.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Heart-Assist Devices*
  • Models, Cardiovascular
  • Prosthesis Design
  • Pulsatile Flow
  • Sheep