Realization of fiber-based laser Doppler vibrometer with serrodyne frequency shifting

Appl Opt. 2011 Jun 10;50(17):2809-14. doi: 10.1364/AO.50.002809.

Abstract

We demonstrate a laser Doppler vibrometer (LDV) based on the serrodyne frequency shifting technique. A proof-of-principle system is implemented on the basis of fiber-optic components but opens the way toward an ultracompact integrated LDV system on a silicon chip. With a low laser power of 50 μW, the serrodyne LDV was able to measure submicrometer vibrations with frequencies in the audio range.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Air
  • Algorithms
  • Equipment Design
  • Fiber Optic Technology / instrumentation*
  • Hearing Loss
  • Laser-Doppler Flowmetry / instrumentation*
  • Laser-Doppler Flowmetry / methods
  • Lasers
  • Pressure
  • Transducers
  • Vibration