Development of a new consumable-free thermal modulator for comprehensive two-dimensional gas chromatography

J Chromatogr A. 2011 May 20;1218(20):3070-9. doi: 10.1016/j.chroma.2011.03.024. Epub 2011 Mar 21.

Abstract

A simple and cost-effective GC×GC modulator requiring no moving parts or consumables, hence suitable for field analysis and monitoring, was developed. The modulator was constructed from a specially designed Silcosteel® trapping capillary, installed outside the GC oven, and coated inside with polydimethylsiloxane (PDMS) stationary phase. Dual-stage modulation was accomplished by resistively heating alternate segments of the trap with a custom-designed capacitive discharge power supply. The performance of the proposed modulator was comparable to many GC×GC systems currently in use, with the injection band widths as low as 60 ms at half height. With proper selection of the stationary phase in the trap, the modulator can be used for the analysis of complex mixtures with volatility range spanning from n-C5 to n-C40.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Alkanes / chemistry
  • Chromatography, Gas / instrumentation*
  • Cold Temperature
  • Equipment Design
  • Flame Ionization
  • Hot Temperature
  • Models, Chemical
  • Volatilization

Substances

  • Alkanes