Two-point dixon method with flexible echo times

Magn Reson Med. 2011 Apr;65(4):994-1004. doi: 10.1002/mrm.22679. Epub 2010 Nov 16.

Abstract

The two-point Dixon method is a proton chemical shift imaging technique that produces separated water-only and fat-only images from a dual-echo acquisition. It is shown how this can be achieved without the usual constraints on the echo times. A signal model considering spectral broadening of the fat peak is proposed for improved water/fat separation. Phase errors, mostly due to static field inhomogeneity, must be removed prior to least-squares estimation of water and fat. To resolve ambiguity of the phase errors, a corresponding global optimization problem is formulated and solved using a message-passing algorithm. It is shown that the noise in the water and fat estimates matches the Cramér-Rao bounds, and feasibility is demonstrated for in vivo abdominal breath-hold imaging. The water-only images were found to offer superior fat suppression compared with conventional spectrally fat suppressed images.

MeSH terms

  • Adipose Tissue / anatomy & histology*
  • Algorithms*
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Information Storage and Retrieval / methods*
  • Liver / anatomy & histology*
  • Magnetic Resonance Imaging / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity