Measurement of ocular aberrations in downward gaze using a modified clinical aberrometer

Biomed Opt Express. 2011 Feb 1;2(3):452-63. doi: 10.1364/BOE.2.000452.

Abstract

Changes in corneal optics have been measured after downward gaze. However, ocular aberrations during downward gaze have not been previously measured. A commercial Shack-Hartmann aberrometer (COAS-HD) was modified by adding a relay lens system and a rotatable beam splitter to allow on-axis aberration measurements in primary gaze and downward gaze with binocular fixation. Measurements with the modified aberrometer (COAS-HD relay system) in primary and downward gaze were validated against a conventional aberrometer. In human eyes, there were significant changes (p<0.05) in defocus C(2,0), primary astigmatism C(2,2) and vertical coma C(3,-1) in downward gaze (25 degrees) compared to primary gaze, indicating the potential influence of biomechanical forces on the optics of the eye in downward gaze. To demonstrate a further clinical application of this modified aberrometer, we measured ocular aberrations when wearing a progressive addition lens (PAL) in primary gaze (0 degree), 15 degrees downward gaze and 25 degrees downward gaze.

Keywords: (330.4460) Ophthalmic optics and devices; (330.5370) Physiological optics; (330.7327) Visual optics, ophthalmic instrumentation.