[Sutureless hepatic transection using a new radiofrequency assisted device. Theoretical model, experimental study and clinic trial]

Cir Esp. 2011 Mar;89(3):145-51. doi: 10.1016/j.ciresp.2010.10.001. Epub 2011 Feb 2.
[Article in Spanish]

Abstract

The ideal instrument for performing hepatic transection should combine safe and rapid haemostasis in a single tool. We present a new multidisciplinary investigation designed to develop a hepatic transection device assisted by radiofrequency (RF); the investigation included: a computerised theoretical model, and experimental study and a clinical trial of this device. The theoretic modelling was performed by computer, based on the Finite Elements Method (FEM), with the objective of studying the distribution of electrical energy and temperature in the tissue, and to assess the effect of the characteristics of the instrument. The experimental study, based on an in vivo porcine model, suggested that the new instrument would allow the transection velocity of the hepatic parenchyma to be increased with lower bleeding per transection area compared with other techniques extensively used in liver surgery. These data should enable the first phase of clinical trial to be conducted, with preliminary results that suggest that the new device is safe and effective.

Publication types

  • Clinical Trial
  • English Abstract
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Animals
  • Catheter Ablation / instrumentation*
  • Female
  • Hepatectomy / methods*
  • Humans
  • Male
  • Middle Aged
  • Models, Theoretical
  • Swine