Gravimetric wear analysis and particulate characterization of bilateral facet-augmentation system--PercuDyn™

Biomed Mater Eng. 2010;20(6):329-38. doi: 10.3233/BME-2010-0646.

Abstract

Dynamic stabilization systems are emerging as an alternative to fusion instrumentation. However, cyclic loading and micro-motion at various interfaces may produce wear debris leading to adverse tissue reactions such as osteolysis. Ten million cycles of wear test was performed for PercuDyn™ in axial rotation and the wear profile and the wear rate was mapped. A validation study was undertaken to assess the efficiency of wear debris collection which accounted for experimental errors. The mean wear debris measured at the end of 10 million cycles was 4.01 mg, based on the worst-case recovery rate of 68.2%. Approximately 40% of the particulates were less than 5 μm; 92% less than 10 μm. About 43% of particulates were spherical in shape, 27% particulates were ellipsoidal and the remaining particles were of irregular shapes. The PercuDyn™ exhibited an average polymeric wear rate of 0.4 mg/million cycles; substantially less than the literature derived studies for other motion preservation devices like the Bryan disc and Charité disc. Wear debris size and shape were also similar to these devices.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Arthroplasty, Replacement / instrumentation*
  • Arthroplasty, Replacement / methods
  • Computer Simulation
  • Equipment Failure Analysis / instrumentation
  • Equipment Failure Analysis / methods*
  • Humans
  • Intervertebral Disc
  • Intervertebral Disc Degeneration / pathology
  • Intervertebral Disc Degeneration / surgery
  • Materials Testing / methods*
  • Models, Biological
  • Osteolysis / etiology
  • Osteolysis / surgery
  • Particle Size
  • Prosthesis Failure / etiology
  • Rotation / adverse effects
  • Spine / physiology
  • Weight-Bearing / physiology