Implementation of a continuous scanning procedure and a line scan camera for thin-sheet laser imaging microscopy

Biomed Opt Express. 2010 Aug 13;1(2):598-609. doi: 10.1364/BOE.1.000598/.

Abstract

We report development of a continuous scanning procedure and the use of a time delay integration (TDI) line scan camera for a light-sheet based microscope called a thin-sheet laser imaging microscope (TSLIM). TSLIM is an optimized version of a light-sheet fluorescent microscope that previously used a start/stop scanning procedure to move the specimen through the thinnest portion of a light-sheet and stitched the image columns together to produce a well-focused composite image. In this paper, hardware and software enhancements to TSLIM are described that allow for dual sided, dual illumination lasers, and continuous scanning of the specimen using either a full-frame CCD camera and a TDI line scan camera. These enhancements provided a ~70% reduction in the time required for composite image generation and a ~63% reduction in photobleaching of the specimen compared to the start/stop procedure.