A dissociative fluorescence enhancement technique for one-step time-resolved immunoassays

Anal Bioanal Chem. 2011 Feb;399(4):1677-82. doi: 10.1007/s00216-010-4485-y. Epub 2010 Dec 15.

Abstract

The limitation of current dissociative fluorescence enhancement techniques is that the lanthanide chelate structures used as molecular probes are not stable enough in one-step assays with high concentrations of complexones or metal ions in the reaction mixture since these substances interfere with lanthanide chelate conjugated to the detector molecule. Lanthanide chelates of diethylenetriaminepentaacetic acid (DTPA) are extremely stable, and we used EuDTPA derivatives conjugated to antibodies as tracers in one-step immunoassays containing high concentrations of complexones or metal ions. Enhancement solutions based on different β-diketones were developed and tested for their fluorescence-enhancing capability in immunoassays with EuDTPA-labelled antibodies. Characteristics tested were fluorescence intensity, analytical sensitivity, kinetics of complex formation and signal stability. Formation of fluorescent complexes is fast (5 min) in the presented enhancement solution with EuDTPA probes withstanding strong complexones (ethylenediaminetetra acetate (EDTA) up to 100 mM) or metal ions (up to 200 μM) in the reaction mixture, the signal is intensive, stable for 4 h and the analytical sensitivity with Eu is 40 fmol/L, Tb 130 fmol/L, Sm 2.1 pmol/L and Dy 8.5 pmol/L. With the improved fluorescence enhancement technique, EDTA and citrate plasma samples as well as samples containing relatively high concentrations of metal ions can be analysed using a one-step immunoassay format also at elevated temperatures. It facilitates four-plexing, is based on one chelate structure for detector molecule labelling and is suitable for immunoassays due to the wide dynamic range and the analytical sensitivity.

MeSH terms

  • Antibodies / chemistry
  • Chelating Agents / chemistry
  • Chorionic Gonadotropin, beta Subunit, Human / blood*
  • Fluorescence*
  • Humans
  • Immunoassay*
  • Lanthanoid Series Elements / chemistry
  • Organometallic Compounds / chemistry
  • Reference Values
  • Spectrometry, Fluorescence
  • Staining and Labeling
  • Time Factors

Substances

  • Antibodies
  • Chelating Agents
  • Chorionic Gonadotropin, beta Subunit, Human
  • Lanthanoid Series Elements
  • Organometallic Compounds