Accuracy of a digital skinfold system for measuring skinfold thickness and estimating body fat

Br J Nutr. 2011 Feb;105(3):478-84. doi: 10.1017/S0007114510003727. Epub 2010 Dec 7.

Abstract

The use of skinfold thickness measurements to evaluate the distribution of subcutaneous adipose tissue and to predict body fat has recognised advantages. However, the different types of skinfold calliper available present limitations that make them unattractive and perhaps less used in daily practice. The purpose of the present study was to evaluate the accuracy and functionality of a new digital skinfold system, the Liposoft 2008+Adipsmeter V0 (LA), for measuring skinfold thickness and determining body fat proportion (%BF). Skinfold thickness measurements made by the LA were compared with those obtained with a Harpenden (H) calliper from two samples of adults (n 45) and older adults (n 56) in a university-based cross-sectional study. A comparison was also conducted between estimated %BF from skinfolds and dual-energy X-ray absorptiometry. Bland and Altman plots show that skinfolds measured by the LA and H calliper are in high agreement, with a mean difference of 0·3 (95% CI -3·1, 3·4) mm. In regard to the %BF estimated from LA and H skinfolds measurement, the LA produced a similar approximation to dual-energy X-ray absorptiometry %BF, with a mean difference of 0·2 (95% CI -0·8, 1·2) %, compared with %BF obtained with the H calliper. The LA system is an accurate instrumentation and represents an innovation in the evaluation of skinfold thickness and body composition based on anthropometric measurement.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorptiometry, Photon / methods
  • Adipose Tissue / anatomy & histology*
  • Adult
  • Aged
  • Anthropometry / instrumentation*
  • Cross-Sectional Studies
  • Female
  • Humans
  • Male
  • Middle Aged
  • Reproducibility of Results
  • Skinfold Thickness*