X-ray crystal spectrometer for opacity measurements in the 8-18 Å spectral range at the LULI laser facility

Rev Sci Instrum. 2010 Oct;81(10):10E327. doi: 10.1063/1.3491285.

Abstract

An x-ray crystal spectrometer was built in order to measure opacities in the 8-18 Å spectral range with an average spectral resolution of <λ/δλ> ∼ 400. It has been successfully used at the LULI-2000 laser facility (See C. Sauteret, rapport LULI 2001, 88 (2002) at École Polytechnique (France) to measure in the same experimental conditions the 2p-3d transitions of several elements with the neighboring atomic number Z: Fe, Ni, Cu, and Ge [G. Loisel et al., High Energy Density Phys. 5, 173 (2009)]. Hence, a spectrometer with a wide spectral range is needed. This spectrometer features two lines of sight. In this example, one line of sight looks through the sample and the other one is looking directly at the backlighter emission. Both are outfitted with a spherical condensing mirror. A TlAP crystal is used for spectral dispersion. Detection is made with an image plate Fuji BAS TR2025, which is sensitive to x rays. We present some experimental results showing the performances of this spectrometer.