Limitations of the DF-4 defibrillator connector necessitating device removal

Pacing Clin Electrophysiol. 2012 Jan;35(1):e24-6. doi: 10.1111/j.1540-8159.2010.02873.x. Epub 2010 Aug 17.

Abstract

The DF-4 implantable defibrillator connector was recently released for clinical practice. This connector facilitates lead to device connection, reduces bulk in the device pocket, and eliminates the risk of incorrect device connection. Unfortunately, new technology often introduces new challenges. We report the case of a 63-year-old male with chronic systolic heart failure referred for cardiac resynchronization therapy-defibrillator implant. Limitations implicit to the current iteration of this technology include a lack of additional connectivity. In the present case, these limitations ultimately warranted device removal and reimplant with a traditional trifurcating IS-1/DF-1 connector.

Publication types

  • Case Reports

MeSH terms

  • Defibrillators, Implantable*
  • Device Removal / methods*
  • Equipment Failure*