Determination of neptunium in environmental samples by extraction chromatography after valence adjustment

Appl Radiat Isot. 2010 Dec;68(12):2132-9. doi: 10.1016/j.apradiso.2010.06.021. Epub 2010 Jul 1.

Abstract

Neptunium(V) ions are unstable in acid media, which limits their extraction on chromatographic resins. We developed a novel analytical method to measure Np by either α-spectrometry or inductively coupled plasma mass spectrometry (ICP-MS) after extraction chromatography as Np(VI). We investigated the reactivity of various oxidizing reagents, and determined the retention capacity of Np(IV, V, and VI) on various extraction chromatographic supports. A simple method using two UTEVA resins was used to rapidly detect Np in soil and sediment samples.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography / methods*
  • Mass Spectrometry
  • Neptunium / analysis*

Substances

  • Neptunium