Analysis of stereotactic accuracy of the cosman-robert-wells frame and nexframe frameless systems in deep brain stimulation surgery

Stereotact Funct Neurosurg. 2010;88(5):288-95. doi: 10.1159/000316761. Epub 2010 Jun 24.

Abstract

Introduction: The primary goal of stereotactic systems in deep brain stimulation (DBS) surgery is accurate delivery of a DBS lead to a target identified on imaging. Thus, it is critical to understand the accuracy of the stereotactic systems and the factors which may be associated with a decrease in accuracy.

Methods: Ninety patients underwent microelectrode recording-guided placement of 139 DBS leads by a single surgeon using the Cosman-Roberts-Wells (CRW) frame (n = 70) or a frameless skull-mounted trajectory guide (Nexframe; n = 69). The final DBS location was identified on a postoperative CT fused to the preoperative CT and MRI scans. The difference between this final location and the expected location was calculated.

Results: The vector error was 2.65 mm (standard error, 0.22) for the frame and 2.78 mm (standard error, 0.25) for the frameless methods (p = 0.69). There was a gradual decline in error for both systems over time, as the vector error of the last 20 implants was 1.99 for the CRW frame and 2.04 for the Nexframe (p = 0.86).

Conclusions: This study shows that the CRW frame and Nexframe frameless systems have equivalent accuracy. Furthermore, the accuracy of both techniques improved over time, from 3 mm initially to 2 mm with current techniques.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Analysis of Variance
  • Chi-Square Distribution
  • Deep Brain Stimulation / instrumentation
  • Globus Pallidus / surgery*
  • Humans
  • Implantable Neurostimulators
  • Male
  • Parkinson Disease / surgery*
  • Stereotaxic Techniques / instrumentation*
  • Subthalamic Nucleus / surgery*
  • Treatment Outcome