Abstract

Background: This multicenter, prospective, randomized, controlled, parallel trial compares the efficacy of biventricular (BIV) versus right ventricular (RV) antitachycardia pacing (ATP) in terminating all kinds of ventricular tachycardia (VT).

Methods: Five hundred twenty-six patients implanted with a cardiac resynchronization therapy defibrillator (CRT-D) device were enrolled and randomized 1:1 to either BIV (266) or RV (260) ATP (single burst 8 pulse, 88% coupling interval) and were followed up for 12 months.

Results: During 12 months' follow-up, 1,077 ventricular episodes in 180 patients were detected and classified: 634 true VTs divided into 69 ventricular fibrillation (VF) (11%), 202 fast ventricular tachycardia (FVT) (32%), and 363 VT (57%). A comparable first ATP efficacy (BIV 65% vs RV 68%, P = .59) was observed in FVT + VT, in VT zone (BIV 62% vs RV 71%, P = .25), and in FVT zone (BIV 71% vs RV 61%, P = .34). A trend toward lower accelerations during ATP applied to FVT was observed in the BIV group (3.5% BIV vs 10.2% RV, P = .163). No syncope/presyncope occurred during ATP for FVT in the BIV group versus 4 events (3.2%) in the RV group (P = .016). biventricular ATP was more effective in treating FVT in coronary artery disease (CAD) patients (P = .032), whereas both modalities presented similar efficacy in patients with non-CAD etiology (P = .549).

Conclusions: Antitachycardia pacing is effective in patients implanted with a CRT-D device. No significant differences in efficacy emerged between BIV- and RV-delivered ATP in the general population, whereas BIV ATP seems to present a safer profile in ischemic patients.

Trial registration: ClinicalTrials.gov NCT00147290.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Cardiac Pacing, Artificial / methods*
  • Defibrillators, Implantable*
  • Disease-Free Survival
  • Electric Countershock / instrumentation*
  • Electrocardiography
  • Female
  • Follow-Up Studies
  • Heart Ventricles / physiopathology*
  • Humans
  • Male
  • Prospective Studies
  • Single-Blind Method
  • Survival Rate
  • Tachycardia, Ventricular / mortality
  • Tachycardia, Ventricular / physiopathology
  • Tachycardia, Ventricular / therapy*
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT00147290