Dual-band water and lipid suppression for MR spectroscopic imaging at 3 Tesla

Magn Reson Med. 2010 Jun;63(6):1486-92. doi: 10.1002/mrm.22324.

Abstract

A dual-band water and lipid suppression sequence was developed for multislice sensitivity-encoded proton MR spectroscopic imaging of the human brain. The presaturation scheme consisted of five dual-band frequency-modulated radiofrequency pulses based on hypergeometric functions integrated with eight outer volume suppression (OVS) pulses. The flip angles of the dual-band pulses were optimized through computer simulations to maximize suppression factors over a range of transmitter amplitude of radiofrequency field and water and lipid T(1) values. The resulting hypergeometric dual band with OVS (HGDB + OVS) sequence was implemented at 3 T in a multislice sensitivity-encoded proton MR spectroscopic imaging experiment and compared to a conventional water suppression scheme (variable pulse power and optimized relaxation delays (VAPOR)) with OVS. The HGDB sequence was significantly shorter than the VAPOR sequence (230 versus 728 msec). Both HGDB + OVS and VAPOR + OVS produced good water suppression, while lipid suppression with the HGDB + OVS sequence was far superior. In sensitivity-encoded proton MR spectroscopic imaging data, artifacts from extracranial lipid signals were significantly lower with HGDB + OVS. The shorter duration of HGDB compared to VAPOR also allows reduced pulse repetition time values in the multislice acquisition.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brain / diagnostic imaging*
  • Humans
  • Image Enhancement / methods*
  • Lipids*
  • Magnetic Resonance Spectroscopy / methods*
  • Phantoms, Imaging
  • Radiography
  • Water*

Substances

  • Lipids
  • Water