Luminance range compression for video film digitizers

Med Phys. 1991 Mar-Apr;18(2):198-205. doi: 10.1118/1.596707.

Abstract

Video cameras are used in many film digitization and teleradiology systems. However, the density range of medical radiographs often exceeds the dynamic range of the camera, and all diagnostic information in the original image may not be captured. Information in both the high and low density areas of the film can be captured in a single video frame if the transmitted luminance range of the radiograph is reduced. This can be accomplished by spatially modulating the back illumination of the film such that areas of lesser density receive less illumination while areas of greater density receive greater illumination. In this work, the use of a video monitor is shown to be an effective means to provide spatially modulated light for compressing the transmitted luminance range and thereby expanding the apparent dynamic range of the video camera. A simple computer-interfaced video feedback system that determines the appropriate compression mask and a scheme for linearization of system response are described. This system provides an interactive means for control of the degree of range compression.

MeSH terms

  • Analog-Digital Conversion*
  • Humans
  • Light
  • Television / instrumentation*