Effect of bone quality on the artificial temporomandibular joint condylar prosthesis

Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010 Jun;109(6):e1-5. doi: 10.1016/j.tripleo.2010.02.005.

Abstract

Objective: This study evaluated the effect of bone quality on the stability of and stress in the artificial temporomandibular joint (TMJ) condylar prosthesis as well as on the stress and strain distributions in the bone.

Study design: Three-dimensional finite element models of the mandible and an artificial TMJ condylar prosthesis using 3 screws were established. Four types of bone were modeled by varying the elastic modulus of cancellous bone.

Results: Increasing the bone quality from type IV to type I reduced the maximum relative micromotion between implant and bone, maximum von Mises stresses in the implant and surrounding cortical bone, and maximum von Mises strain in the surrounding cancellous bone by 17.5%, 10.5%, 12.0%, and 39.4%, respectively.

Conclusion: The bone quality of the mandible has only a minor effect on the stability of and stress in an artificial TMJ implant and cortical bone, but it greatly affects the microstrain in cancellous bone.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biomechanical Phenomena
  • Bone Density*
  • Computer Simulation
  • Dental Prosthesis Design
  • Dental Stress Analysis*
  • Elastic Modulus
  • Finite Element Analysis
  • Humans
  • Imaging, Three-Dimensional
  • Joint Prosthesis*
  • Mandibular Condyle / pathology*
  • Models, Anatomic*
  • Stress, Mechanical
  • Temporomandibular Joint / pathology*