Phase and amplitude correction for multi-echo water-fat separation with bipolar acquisitions

J Magn Reson Imaging. 2010 May;31(5):1264-71. doi: 10.1002/jmri.22111.

Abstract

Purpose: To address phase and amplitude errors for multi-point water-fat separation with "bipolar" acquisitions, which efficiently collect all echoes with alternating read-out gradient polarities in one repetition.

Materials and methods: With the bipolar acquisitions, eddy currents and other system nonidealities can induce inconsistent phase errors between echoes, disrupting water-fat separation. Previous studies have addressed phase correction in the read-out direction. However, the bipolar acquisitions may be subject to spatially high order phase errors as well as an amplitude modulation in the read-out direction. A method to correct for the 2D phase and amplitude errors is introduced. Low resolution reference data with reversed gradient polarities are collected. From the pair of low-resolution data collected with opposite gradient polarities, the two-dimensional phase errors are estimated and corrected. The pair of data are then combined for water-fat separation.

Results: We demonstrate that the proposed method can effectively remove the high order errors with phantom and in vivo experiments, including obliquely oriented scans.

Conclusion: For bipolar multi-echo acquisitions, uniform water-fat separation can be achieved by removing high order phase errors with the proposed method.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adipose Tissue / anatomy & histology*
  • Algorithms
  • Artifacts*
  • Body Water / cytology*
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Imaging, Three-Dimensional / methods
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Subtraction Technique*