Performance of a novel repositioning head frame for gamma knife perfexion and image-guided linac-based intracranial stereotactic radiotherapy

Int J Radiat Oncol Biol Phys. 2010 Sep 1;78(1):306-13. doi: 10.1016/j.ijrobp.2009.11.001. Epub 2010 Apr 10.

Abstract

Purpose: To evaluate the geometric positioning and immobilization performance of a vacuum bite-block repositioning head frame (RHF) system for Perfexion (PFX-SRT) and linac-based intracranial image-guided stereotactic radiotherapy (SRT).

Methods and materials: Patients with intracranial tumors received linac-based image-guided SRT using the RHF for setup and immobilization. Three hundred thirty-three fractions of radiation were delivered in 12 patients. The accuracy of the RHF was estimated for linac-based SRT with online cone-beam CT (CBCT) and for PFX-SRT with a repositioning check tool (RCT) and offline CBCT. The RCT's ability to act as a surrogate for anatomic position was estimated through comparison to CBCT image matching. Immobilization performance was evaluated daily with pre- and postdose delivery CBCT scans and RCT measurements.

Results: The correlation coefficient between RCT- and CBCT-reported displacements was 0.59, 0.75, 0.79 (Right, Superior, and Anterior, respectively). For image-guided linac-based SRT, the mean three-dimensional (3D) setup error was 0.8 mm with interpatient (Sigma) and interfraction (sigma) variations of 0.1 and 0.4 mm, respectively. For PFX-SRT, the initial, uncorrected mean 3D positioning displacement in stereotactic coordinates was 2.0 mm, with Sigma = 1.1 mm and sigma = 0.8 mm. Considering only RCT setups <1mm (PFX action level) the mean 3D positioning displacement reduced to 1.3 mm, with Sigma = 0.9 mm and sigma = 0.4 mm. The largest contributing systematic uncertainty was in the superior-inferior direction (mean displacement = -0.5 mm; Sigma = 0.9 mm). The largest mean rotation was 0.6 degrees in pitch. The mean 3D intrafraction motion was 0.4 +/- 0.3 mm.

Conclusion: The RHF provides excellent immobilization for intracranial SRT and PFX-SRT. Some small systematic uncertainties in stereotactic positioning exist and must be considered when generating PFX-SRT treatment plans. The RCT provides reasonable surrogacy for internal anatomic displacement.

Publication types

  • Evaluation Study

MeSH terms

  • Brain Neoplasms / surgery*
  • Cone-Beam Computed Tomography / methods
  • Cranial Irradiation / instrumentation*
  • Dose Fractionation, Radiation
  • Head
  • Humans
  • Immobilization / instrumentation*
  • Immobilization / methods
  • Medical Illustration
  • Patient Positioning / instrumentation*
  • Prospective Studies
  • Radiosurgery / instrumentation*
  • Radiosurgery / methods
  • Radiotherapy, Computer-Assisted / instrumentation
  • Radiotherapy, Computer-Assisted / methods
  • Vacuum