Technical aspects of MR perfusion

Eur J Radiol. 2010 Dec;76(3):304-13. doi: 10.1016/j.ejrad.2010.02.017. Epub 2010 Apr 3.

Abstract

The most common methods for measuring perfusion with MRI are arterial spin labelling (ASL), dynamic susceptibility contrast (DSC-MRI), and T(1)-weighted dynamic contrast enhancement (DCE-MRI). This review focuses on the latter approach, which is by far the most common in the body and produces measures of capillary permeability as well. The aim is to present a concise but complete overview of the technical issues involved in DCE-MRI data acquisition and analysis. For details the reader is referred to the references. The presentation of the topic is essentially generic and focuses on technical aspects that are common to all DCE-MRI measurements. For organ-specific problems and illustrations, we refer to the other papers in this issue. In Section 1 "Theory" the basic quantities are defined, and the physical mechanisms are presented that provide a relation between the hemodynamic parameters and the DCE-MRI signal. Section 2 "Data acquisition" discusses the issues involved in the design of an optimal measurement protocol. Section 3 "Data analysis" summarizes the steps that need to be taken to determine the hemodynamic parameters from the measured data.

Publication types

  • Review

MeSH terms

  • Contrast Media*
  • Humans
  • Image Enhancement / methods*
  • Magnetic Resonance Imaging / methods*
  • Permeability
  • Signal Processing, Computer-Assisted