MONICA: a compact, portable dual gamma camera system for mouse whole-body imaging

Nucl Med Biol. 2010 Apr;37(3):245-53. doi: 10.1016/j.nucmedbio.2009.12.003.

Abstract

Introduction: We describe a compact, portable dual-gamma camera system (named "MONICA" for MObile Nuclear Imaging CAmeras) for visualizing and analyzing the whole-body biodistribution of putative diagnostic and therapeutic single photon emitting radiotracers in animals the size of mice.

Methods: Two identical, miniature pixelated NaI(Tl) gamma cameras were fabricated and installed "looking up" through the tabletop of a compact portable cart. Mice are placed directly on the tabletop for imaging. Camera imaging performance was evaluated with phantoms and field performance was evaluated in a weeklong In-111 imaging study performed in a mouse tumor xenograft model.

Results: Tc-99m performance measurements, using a photopeak energy window of 140 keV+/-10%, yielded the following results: spatial resolution (FWHM at 1 cm), 2.2 mm; sensitivity, 149 cps (counts per seconds)/MBq (5.5 cps/microCi); energy resolution (FWHM, full width at half maximum), 10.8%; count rate linearity (count rate vs. activity), r(2)=0.99 for 0-185 MBq (0-5 mCi) in the field of view (FOV); spatial uniformity, <3% count rate variation across the FOV. Tumor and whole-body distributions of the In-111 agent were well visualized in all animals in 5-min images acquired throughout the 168-h study period.

Conclusion: Performance measurements indicate that MONICA is well suited to whole-body single photon mouse imaging. The field study suggests that inter-device communications and user-oriented interfaces included in the MONICA design facilitate use of the system in practice. We believe that MONICA may be particularly useful early in the (cancer) drug development cycle where basic whole-body biodistribution data can direct future development of the agent under study and where logistical factors, e.g., limited imaging space, portability and, potentially, cost are important.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Animals
  • Equipment Design
  • Equipment Failure Analysis
  • Gamma Cameras / veterinary*
  • Mice
  • Miniaturization
  • Radionuclide Imaging / instrumentation*
  • Radionuclide Imaging / veterinary*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Whole Body Imaging / instrumentation*
  • Whole Body Imaging / veterinary*