End-to-end differential contactless conductivity sensor for microchip capillary electrophoresis

Anal Chem. 2010 Apr 15;82(8):3270-5. doi: 10.1021/ac100041p.

Abstract

In this contribution, a novel measurement approach for miniaturized capillary electrophoresis (CE) devices is presented: End-to-end differential capacitively coupled contactless conductivity measurement. This measurement technique is applied to a miniaturized CE device fabricated in low-temperature cofired ceramics (LTCC) multilayer technology. The working principle is based on the placement of two distinct detector areas near both ends of the fluid inlet and outlet of the separation channel. Both output signals are subtracted from each other, and the resulting differential signal is amplified and measured. This measurement approach has several advantages over established, single-end detectors: The high baseline level resulting from parasitic stray capacitance and buffer conductivity is reduced, leading to better signal-to-noise ratio and hence higher measurement sensitivity. Furthermore, temperature and, thus, baseline drift effects are diminished owing to the differentiating nature of the system. By comparing the peak widths measured with both detectors, valuable information about zone dispersion effects arising during the separation is obtained. Additionally, the novel measurement scheme allows the determination of dispersion effects that occur at the time of sample injection. Optical means of dispersion evaluation are ineffective because of the opaque LTCC substrate. Electrophoretic separation experiments of inorganic ions show sensitivity enhancements by about a factor of 30-60 compared to the single-end measurement scheme.