False elevation of blood glucose levels measured by GDH-PQQ-based glucometers occurs during all daily dwells in peritoneal dialysis patients using icodextrin

Perit Dial Int. 2010 May-Jun;30(3):329-35. doi: 10.3747/pdi.2008.00285. Epub 2010 Feb 26.

Abstract

Objective: False elevation of blood glucose levels measured by glucose dehydrogenase pyrroloquinoline quinone (GDH-PQQ)-based glucose self-monitoring systems; glucometer) in peritoneal dialysis (PD) patients using icodextrin solution has been well documented. However, adverse hypoglycemic events caused by misreadings for blood glucose are still being reported. We aimed to study blood glucose levels measured simultaneously using different methods in PD patients with switching of icodextrin, and throughout daily exchanges either using icodextrin or not.

Design: We recruited 100 PD patients, including 40 using icodextrin; 128 hemodialysis patients served as a reference. Fasting serum glucose was measured using our laboratory reference method (LAB) and 2 glucose self-monitoring systems based on glucose dehydrogenase nicotinamide adenine dinucleotide (GDH-NAD) and GDH-PQQ respectively. 80 PD patients had a second follow-up study. A time course study was performed in 16 PD patients through measuring fingertip glucose using the 2 glucose self-monitoring systems during daily exchanges.

Result: The differences in measured serum glucose levels in (PQQ minus LAB) versus (NAD minus LAB) were markedly increased in PD patients using icodextrin compared to other patient groups, and was further confirmed by the follow-up study in patients that switched to icodextrin. The high serum glucose levels measured by the GDH-PQQ-based glucose self-monitoring system were present throughout all exchanges during the day in patients using icodextrin solution.

Conclusion: False elevation of blood glucose measured by GDH-PQQ-based glucose self-monitoring systems exists in patients using icodextrin. To avoid misinterpretation of hyperglycemia and subsequent over-injection of insulin, GDH-PQQ-based glucose self-monitoring systems should not be used in PD patients.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Blood Glucose / analysis*
  • Blood Glucose Self-Monitoring / instrumentation
  • Blood Glucose Self-Monitoring / methods
  • Dialysis Solutions / adverse effects*
  • False Positive Reactions
  • Female
  • Glucans / adverse effects*
  • Glucose / adverse effects*
  • Humans
  • Hyperglycemia / blood
  • Icodextrin
  • Male
  • Middle Aged
  • Peritoneal Dialysis*
  • Renal Dialysis
  • Taiwan