Abstract

Five adrenolytic drugs have been analyzed by liquid chromatography-mass spectrometry (LC-MS). Samples were prepared by solid-phase microextraction (SPME) using polypyrrole fibers coated on stainless steel support as an adsorbent for the drugs. Adsorption efficiencies were 95% and were close for all the drugs investigated. Relative standard deviations (RSD), calculated for samples prepared in standard solutions, were in the range 2.5-13%, however RSD values for the drugs in human plasma were 2.5-4.5%. Using LC-MS the limit of detection (LOD) and the limit of quantification (LOQ) were in the ranges 0.11-0.18 and 0.39-0.54 ng mL(-1), respectively, for the five drugs.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adrenergic Antagonists / blood*
  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Limit of Detection
  • Solid Phase Microextraction / methods*
  • Spectrometry, Mass, Electrospray Ionization / methods*

Substances

  • Adrenergic Antagonists