Implementation of a new low cost detector for low intensity light pulses produced by a radioactive gamma-source

Appl Radiat Isot. 2010 Jul-Aug;68(7-8):1320-3; discussion 1323. doi: 10.1016/j.apradiso.2010.01.011. Epub 2010 Jan 13.

Abstract

The readout of a cheap scintillating fibre was implemented by means of a new solid-state device (Silicon Photomultiplier, SiPM), able to detect very tiny light pulses. The GEANT3.21 code was adopted to simulate the counting detection efficiency and the energy deposited inside the scintillating fibre due to the radiation of a radioactive gamma source. The results obtained show the capability to detect gamma rays producing as few as 3-4 photoelectrons. The new devices may have many applications in radioactivity metrology.