Thermoluminescent chip detector for in vivo dosimetry in pelvis and head & neck cancer treatment

Appl Radiat Isot. 2010 Apr-May;68(4-5):795-8. doi: 10.1016/j.apradiso.2009.12.027. Epub 2009 Dec 21.

Abstract

Our aim is to show the TL dosimetry as a confident QA method for radiotherapy treatments. Before in vivo entrance dose measurements using TLD-100 chips, ECLIPSE TPS-simulated treatments for a Rando anthropomorphic phantom, two for pelvis and one head & neck. In Vivo measurements results with (60)Co beam remained within +/-5% limits. Results for 6 and 15 MV are in conclusion. This is a National Cancer Institute/RJ/Brazil study under the 13.111-IAEA Coordinated Research Project.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis
  • Head and Neck Neoplasms / radiotherapy*
  • Humans
  • Pelvic Neoplasms / radiotherapy*
  • Radiotherapy Dosage
  • Radiotherapy, Conformal / instrumentation*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Thermoluminescent Dosimetry / instrumentation*
  • Transducers*