Evaluation of multicoil breast arrays for parallel imaging

J Magn Reson Imaging. 2010 Feb;31(2):328-38. doi: 10.1002/jmri.22023.

Abstract

Purpose: To evaluate three multicoil breast arrays for both conventional and SENSE-accelerated imaging.

Materials and methods: Two commercially available 8-element coils and a prototype 16-element coil were compared. One 8-element array had adjustable coils located next to the breast tissue and the other had a fixed coil arrangement; both were designed to allow parallel imaging in the left-right direction. The 16-element coil was designed to have coil sensitivity variation in both the left-right and superior-inferior directions, and also had adjustable coils. Their performance was assessed in terms of signal-to-noise ratio (SNR), g-factor, and uniformity with a custom-built phantom.

Results: The 16-element array with adjustable coils provided the highest SNR, while the 8-element coil with a fixed coil arrangement had the best uniformity. All coils performed well for SENSE acceleration in the left-right direction. The 8-element coils did not have the capability for acceleration in the superior-inferior direction across the whole volume. The 16-element coil enabled acceleration in the superior-inferior direction in addition to the left-right direction.

Conclusion: Smaller, adjustable coil elements located next to breast tissue can provide greater SNR than larger, fixed coil elements. A multicoil breast array with high intrinsic SNR and low g-factors enables high-quality parallel imaging.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Breast / anatomy & histology*
  • Equipment Design
  • Equipment Failure Analysis
  • Female
  • Humans
  • Image Enhancement / instrumentation*
  • Magnetic Resonance Imaging / instrumentation*
  • Magnetics / instrumentation*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Transducers*