Hand-assisted laparoscopic partial nephrectomy: a controlled study of bipolar inline radiofrequency ablation device in Swine

Surg Innov. 2010 Mar;17(1):57-62. doi: 10.1177/1553350609360354. Epub 2010 Jan 21.

Abstract

Background: Bipolar inline radiofrequency ablation (ILRFA) is an effective way to control bleeding during parenchymal organ transection. This was investigated in a study of laparoscopic hand-assisted partial nephrectomy in swine.

Methods: Nine Landrace pigs were divided between 2 groups; diathermy was employed in the control group for parenchymal transaction and ILRFA was applied in the experimental group through a hand-port and deployed into the resection plane. After complete coagulation, resection was performed using scissors.

Results: The mean intraoperative blood loss was 32 +/- 15 mL in the ILRFA group and 187 +/- 69 mL in the control group: a 82.9% reduction ( P = .015). The mean blood loss per resection area was 2.53 +/- 0.92 mL/cm(2) in the ILRFA group compared with 17.31 +/- 9.05 mL/cm( 2) in controls; the reduction was 85.4% (P = .005).

Conclusions: ILRFA is effective in achieving reducing blood loss and provides a drier operative field for precise dissection.

MeSH terms

  • Animals
  • Catheter Ablation / instrumentation*
  • Electrocoagulation
  • Laparoscopy
  • Models, Animal
  • Nephrectomy / methods*
  • Swine